红外耳温计测量结果的不确定度评定

2024-06-27 23:42:25王静
品牌与标准化 2024年3期
关键词:不确定度温度

王静

【摘要】红外耳温计相对于传统的玻璃体温计、医用电子体温计等,具有测量过程时间短、使用方便和非接触性等优点,在医疗卫生行业、商场等人流量密集场所和个人家庭中都有广泛使用。本文主要对红外耳温计测量结果示值误差的不确定度进行评定。

【关键词】红外耳温计;不确定度;温度

【DOI编码】10.3969/j.issn.1674-4977.2024.03.079

Evaluation of Uncertainty in the Measurement Results of Infrared Ear Thermometers

WANG Jing

(Hainan Provincial Institute of Metrology and Testing, Haikou 570206, China)

Abstract: Compared to traditional glass thermometers and medical electronic thermometers, infrared ear thermometers have the advantages of short measurement time, fast and convenient measurement process. They are widely used in places with high traffic such as the medical and health industry, shopping malls, and personal households. This article evaluates the uncertainty of the indication error in the measurement results of infrared ear thermometers.

Keywords: infrared ear thermometer; uncertainty; temperature

0引言

1概述

1.1测量方法

依据JJG 1164—2019《红外耳温计》检定规程,采用比较法,在校准模式下用被测红外耳温计测量黑体辐射源,所测值与标准铂电阻示值进行比较计算。

1.2环境条件

温度为20℃,相对湿度为65%。检定环境无明显机械振动和强电磁干扰,避免强烈辐射和对流对检定装置和被检耳温计产生干扰,避免剧烈波动。

1.3测量标准

1)标准铂电阻温度计,二等;

2)黑体辐射源:U = 0.1℃,k = 2;

3)制冷恒温槽:温度范围-10~90℃;

4)电测设备:U = 0.1℃,k = 2。

1.4测量对象

测量范围为35.0~41.5℃的红外耳温计。

1.5测量过程

1)检定前先将红外耳温计放置于检定环境中至少30 min。

2)用比较法将标准铂电阻温度计插入恒温槽内并连接好电测设备,黑体辐射源放在恒温槽上,调整恒温槽温度,将恒温槽温度恒定在被检温度点上,稳定后耳温计黑体温度与检定点的偏差不超过0.1℃。

3)标准铂电阻示值稳定后,把红外耳温计调到测试或者校准模式,对着黑体空腔测量红外耳温计温度,分别读取标准铂电阻温度计和红外耳温计的示值。

4)测量时注意耳温计探头前端与空腔开口位置齐平且伸入空腔开口不超过2 mm。耳温计瞄准操作至开始测量之间的时间间隔不超过5 s。对于要求使用探头保护罩的耳温计,检定时应使用探头保护罩,且每次测量更换新的探头保护罩。

1.6检定注意事项

在使用红外耳温测量的过程中有以下几个方面需要注意:

1)探头插入耳朵的深度要达到规定的要求。

2)测量过程中通过拉伸耳朵,保证耳道径直以减少误差。

3)在使用红外耳温计测量前最好进行必要的耳道清理。

2建立数学模型

3方差和传播系数

4不确定度评定

4.1黑体辐射源准确度引入的不确定度u1

黑体辐射源检定证书扩展不确定度为:

5标准不确定度汇总