【摘要】气体稀释装置在各类气体检测报警器、气体分析仪上的应用越来越广泛。通过标准气体和气体稀释装置对同一台一氧化碳红外气体分析仪在选定的校准点进行校准,结合示值误差、示值重复性等校准数据分析了两种校准方式的差异,为相关计量校准工作提供参考。
【关键词】标准气体;气体稀释装置;比较
【DOI编码】10.3969/j.issn.1674-4977.2023.02.019
Comparison of Application of Standard Gas and Gas Dilution Device
LIU Xueen
(Liaocheng Inspection and Testing Center,Liaocheng 252000,China)
Abstract:Gas dilution device is more and more widely used in various types of gas detection alarm and gas analyzer. In this paper,the same carbon monoxide infrared gas analyzer is calibrated at the selected calibration point through standard gas and gas dilution device. Combined with the calibration data such as indication error and indication repeatability,the differences between the two calibration methods are analyzed,which provides a reference for metrological colleagues in practical work in the future.
Key words:standard gas;gas dilution device;comparison
随着我国工业化进程的不断深入,我国对环境保护投入的力度不断加强,针对大气及生产、工作环境中的气体进行检测、分析、报警的各类仪器仪表得到了广泛的应用。GB 30871—2022《危险化学品企业特殊作业安全规范》[1]中多次提到,在作业中应配备便携式气体检测报警仪等设备对可燃气体、有毒有害气体及氧气进行浓度监测。GB 3095—2012《环境空气质量标准》[2]中对环境空气中氮氧化物、二氧化硫、一氧化碳等污染物的浓度限值及测定方法作了明确的规定。GB 13271—2014《锅炉大气污染物排放標准》[3]中对锅炉等固定污染源排放的各类污染物采集、分析、监测等作了详细的要求。由于生产厂家生产的气体分析仪器量程不相同,并且便携式与固定式配合使用,使用钢瓶标准气体对气体分析仪进行检定校准的传统方法已经不能满足要求。因此,气体动态稀释装置在检定校准气体分析仪中得到广泛应用。
气体动态稀释装置可以适用不同种类的标准气体,还可以按照稀释范围灵活选择标准气体浓度值,满足不同生产厂家不同量程传感器的检定校准点的要求,极大地方便了日常的检定校准工作。气体动态稀释装置的配气不确定度可以控制在1%以内,流量的线性误差可以控制在±0.3%以内,气体稀释范围在2%~96%,可以满足大多数气体分析仪器的检定校准。本文对检定校准不同量程的气体分析仪器时,如何选择“源”标准气体,以及“源”标准气体的选择对检定校准结果的影响进行相关试验。
1.1被测对象(见表1)
1.2配套设备
多组分动态配气系统,型号:MF-5B,编号:180901,生产厂家:中科环仪(北京)计量技术有限公司。
1.3标准气体(见表2)
2.1仪器零点及满量程校准
3.1不同浓度测量结果误差
按照公式(1)计算6次测量的示值误差为-0.30%FS;按照公式(2)计算6次测量的示值重复性为0.21%。