【摘要】血液透析机是模拟人体肾功能,纠正体内电解质与维持酸碱度平衡的一组体外平衡装置,其相关参数测量结果的质量控制直接关系到患者的身体健康。本文通过GUM法对血液透析机电导率示值误差、温度示值误差、静脉压示值误差、流量示值误差的测量结果进行不确定度来源分析,并利用贝塞尔公式法、极差法和经验法等对相关参数的测量结果进行不确定度评定。
【关键词】血液透析机;测量结果;不确定度
【DOI编码】10.3969/j.issn.1674-4977.2023.03.013
Uncertainty Analysis of Measurement Results of Hemodialysis Machine Related Parameters
ZHU Yuling
(Tongren Inspection and Testing Institute,Tongren 554300,China)
Abstract:Hemodialysis machines are a group of external balance devices that simulate human renal function,correct electrolytes in the body,and maintain acid-base balance. The quality control of the measurement results of their related parameters is directly related to the life and health of patients. This article uses the GUM method to analyze the sources of uncertainty in the measurement results of electrical conductivity indication error,temperature indication error,venous pressure indication error,and flow indication error of hemodialysis machines,and uses the Bessel formula method to analyze the sources of uncertainty. The range method and empirical method are used to evaluate the uncertainty.
Key words:hemodialysis machine;measurement results;uncertainty
本实验校准依据为JJF 1353—2012《血液透析装置校准规范》,现场校准环境温度为22~25℃,相对湿度小于70%,大气压力为93~100 kPa,校准方法为直接测量法。
因各项互不相关,且式中各分量相互独立,所以灵敏度系数ci = 1,合成方差为各不确定度分量的平方和。
4.1不确定度分量来源分析及计算
4.1.1被校装置引入的不确定度分量ux
1)被校装置透析液电导率示值重复性引入的标准不确定度分量ux1的评定
在符合校准的环境条件下,用校准器具对设定好透析液电导率值的透析机进行10次独立重复实验,被校装置示值如下(单位:mS/cm):14.0、14.0、14.1、14.1、14.1、14.0、14.0、14.0、14.1、14.0,则实验标准偏差为0.05 mS/cm,相对标准偏差为0.36%,选取3次示值的平均值作为被校装置电导率示值,则:
5.1不确定度分量来源分析及计算
5.1.1被校装置引入的不确定度分量ux
1)被校装置显示温度重复性引入的标准不确定度分量ux1的评定
5.4扩展不确定度评定
取k=2,则扩展不确定度为U = kuc = 2×0.05= 0.1℃。
6.1不确定度分量来源分析及计算
6.1.1校准器具测量重复性引入的标准不确定度分量评定
7.1不确定度分量来源分析及计算
7.1.1被校装置引起的不确定度分量
1)被校装置显示流量重复性引入的不确定度分量u01
【参考文献】
[1]曲志勇,王宪民,胡生伟.血液透析装置流量监控示值误差的测量不确定度评定[J].计量与测试技术,2014,41(5):70+72.
[2]李庆芝.血液透析装置透析液流量监控示值误差测量不确定度评定[J].计量与测试技术,2018,45(4):111-112.